High Throughput Analysis of Tetracycline (TCs) in Animal Tissue

2022-05-28 21:49:08 By : Ms. Lois Fang

We use cookies to enhance your experience. By continuing to browse this site you agree to our use of cookies. More info.

Tetracycline antibiotics (TCs) are among the most commonly utilized antibiotics in the animal production sector. They are employed to avoid infection in animals, treat diseases and safeguard animal growth.

Overusing antibiotics in animal-originated food products (for example, milk and meat) may lead to residue violations, resulting in risks to consumers’ health. Tetracycline residues present in food products may lead to a range of problems, including:

To reduce the risks and safeguard consumer health, regulatory authorities worldwide have set out guidelines for acceptable levels of veterinary drugs in food products. In the US, these levels are called ‘tolerances,’ while China, Canada and the European Union refer to them as ‘maximum residue levels’ (MRLs).1,2

Utilizing techniques that are highly sensitive and selective to provide accurate monitoring of tetracycline levels as specified regulatory requirements is a key element in protecting consumer health.

This article outlines a UHPLC/MS/MS method that offers the speed and sensitivity required for carrying out a quantitative analysis of tetracyclines in meat products.

To improve the testing lab’s efficiency, a QS-Works high-throughput autosampler was used for the following reasons:

The method outlined showed excellent sensitivity, recovery, precision, linearity and selectivity needed to analyze tetracyclines in meat products at the low tolerance levels set out by various regulatory authorities around the globe.

A range of hardware was utilized to carry out the analyses. Chromatographic separation was performed on a PerkinElmer QSight® LX50 UHPLC system. The detection process was completed using a PerkinElmer QSight® 420 MS/MS detector.

A QS Works autosampler with a high capacity was utilized to increase the lab’s productivity. The Simplicity 3Q™ software platform was employed to carry out data acquisition and processing and all instrument control.

The MS source parameters and LC method can be seen in Table 1. Multiple reaction monitoring mode (MRM) transitions of the tetracyclines are displayed in Table 2.

Table 1. LC Method and MS Source Parameters. Source: PerkinElmer Food Safety and Quality

Table 2. MRM Transitions. Source: PerkinElmer Food Safety and Quality

A minimum of two MRM transitions were monitored for each analyte. This was to lower the number of false positives and negatives in the outlined method.

The optimizing of MS/MS parameters, which included the choice of parent ions and product ions, entrance voltages (EV/CCL2) and collision energies (CE) was done by infusion of standards.

The optimization of source conditions was achieved via t-infusion of neat standards with LC flow, with the acquisition MS method established based on those optimized conditions.

All reagents, solvents and diluents utilized were LC/MS grade. All tetracycline standards were sourced from Sigma-Aldrich® Inc. These were stored at - 20 °C in a freezer to prevent their degradation.

All stock and mixed drug solutions utilized in the spiking and calibration procedures for all the tetracyclines were prepared in methanol. To decrease the risks of degradation of standards, all working and stock standards were stored in a freezer until needed.

A homogenized sample for use in the procedure is created as follows:

This article aims to evaluate the performance of the high-capacity QS-Works autosampler with a validated sample preparation procedure for the determination of tetracyclines in a meat sample.

Utilizing a high throughput autosampler, for example, the QS-Works, increases testing lab productivity and efficiencies.

The QS-Works allows food testing labs with a large number of samples to process samples more rapidly while giving enhanced data quality and consistency due to its robotic sampling unit (Figure 1).

Figure 1. QS-Works autosampler. Image Credit: PerkinElmer Food Safety and Quality

Reference target levels were determined by utilizing the maximum residue levels (MRLs) established for tetracyclines in foods by Health Canada or tolerances by the US FDA (X in Table 3).

Table 3. Results of retention time, matrix effect (ME), linearity, recovery, and reproducibility (%RSD). Source: PerkinElmer Food Safety and Quality

The detection threshold, or ‘yes/no’ screening level, should be at or below 0.5X. The chromatograms (XIC) of four tetracyclines demonstrated good peak resolution (Figure 2).

Figure 2. XIC chromatograms of a) oxytetracycline, b) tetracycline, c) chlortetracycline, d) doxytetracycline. Image Credit: PerkinElmer Food Safety and Quality

Two ion transitions were observed for confirmation and identification purposes, resulting in greater selectivity and specificity in identification. Every ion ratio was determined by calculating the peak area for less intense ion to peak area for more intense ion in order to generate ion ratios.

Among the main concerns for LC/MS/MS method development are sample matrix effects (MEs). This is especially the case when analyzing food due to the complex and diverse nature of food sample matrices.

ESI is vulnerable to ionization suppression of analytes where charge-competing matrix components are present. Matrix-induced enhancement effects can occur in ESI, which may lead to a large bias in quantification.

To surmount sample matrix effects, a number of approaches have been utilized, including:

Sample matrix effects were calculated by the comparison of the slopes of calibration curves acquired from the meat sample matrix to slopes acquired from reagent-only (RO).

The sample matrix effect (expressed as a percentage) for each analyte was then determined by the percentage difference between the slopes being compared.

The matrix effects for most of the compounds were less than 20% (Table 3).

Matrix-matched calibrations were utilized in this analysis for the quantification of all analytes. This was to control matrix effects and decrease the level of variations in analytical results.

Calibration was carried out utilizing both reagents-only (RO) and matrix-matched (MM) standards.

Every calibration curve that was constructed from both RO and meat sample matrix demonstrated good linearity with a correlation coefficient (R2) larger than 0.99.  For typical examples of calibration curves, see Figures 3 and 4.

Figure 3. Calibration curves of tetracyclines obtained from standards prepared in reagent only. Image Credit: PerkinElmer Food Safety and Quality

Figure 4. Calibration curves of tetracyclines obtained from standards prepared in meat sample matrix. Image Credit: PerkinElmer Food Safety and Quality

The degree of carryover was evaluated by injecting the reagent blank after a high concentration standard. No carryover was noted in any of the tests.

Oxalic acid was added into the aqueous mobile phase3 to prevent chelation of tetracyclines by metal ions of LC parts or column body. This also helped sustain good peak shape and reproducibility over the run.

The method displays good precision with RSD at less than 5% for five replicates. For the effect of using oxalic acid in mobile phase and reconstitute solvent on the reproducibility, see Table 3.

The reproducibility and peak shape of all analytes was improved by the addition of oxalic acid into mobile phase and reconstitute solvent. The limits of quantification (LOQ) for each tetracycline analyzed were lower than the permitted tolerance level in meat samples.

This indicates that the technique meets the levels of sensitivity needed to analyze tolerance levels for tetracycline in meat.

The LOQs for the tetracyclines were estimated based on signal to noise (S/N) ratio of 10. Every LOQ for the drugs studied was 1.5 ± 1.0 µg/L.

These results show this method can be employed for the rapid screening and quantification of tetracyclines in meat samples, while the absolute recoveries of tetracyclines were in the range of 70-120% (Table 3). Utilization of a deuterated tetracycline can compensate for losses of tetracyclines with this extraction process.

A PerkinElmer high throughput QS-Works autosampler and LX 50 UHPLC coupled to a QSight 420 MS/MS system were utilized to demonstrate a UHPLC/MS/MS procedure.

The instrument displays good sensitivity for identifying and quantifying tetracyclines in a homogenized meat sample. The low detection levels attained in the method can help support low regulatory restrictions for everyday screening and quantitation analysis.

Produced from materials originally authored by Saba Hariri, Abir Khaled, Avinash Dalmia, and Feng Qin from PerkinElmer, Inc.

This information has been sourced, reviewed and adapted from materials provided by PerkinElmer Food Safety and Quality.

For more information on this source, please visit PerkinElmer Food Safety and Quality.

Please use one of the following formats to cite this article in your essay, paper or report:

PerkinElmer Food Safety and Quality. (2021, August 11). High Throughput Analysis of Tetracycline (TCs) in Animal Tissue. AZoM. Retrieved on May 28, 2022 from https://www.azom.com/article.aspx?ArticleID=20601.

PerkinElmer Food Safety and Quality. "High Throughput Analysis of Tetracycline (TCs) in Animal Tissue". AZoM. 28 May 2022. <https://www.azom.com/article.aspx?ArticleID=20601>.

PerkinElmer Food Safety and Quality. "High Throughput Analysis of Tetracycline (TCs) in Animal Tissue". AZoM. https://www.azom.com/article.aspx?ArticleID=20601. (accessed May 28, 2022).

PerkinElmer Food Safety and Quality. 2021. High Throughput Analysis of Tetracycline (TCs) in Animal Tissue. AZoM, viewed 28 May 2022, https://www.azom.com/article.aspx?ArticleID=20601.

Do you have a question you'd like to ask regarding this article?

In this interview, AZoM talks to Chris Aylott, Senior Engineer at Newcastle University, about fatigue testing in the development of low carbon transport solutions.

AZoM speaks with Dr. Nicola Ferralis from MIT about his research that has developed a low-cost process of creating carbon fibers from hydrocarbon pitch. This research could lead to the large-scale use of carbon fiber composites in industries that have thus far been limited.

In this interview, AZoM speaks with Marco Enger, Senior Tribologist from GGB, to discuss how nano fillers affect transfer films within tribological systems.

This product profile outlines the Versa Flex Checkweighers from Thermo Fisher Scientific.

The LQA 300 FT-IR wine analyzer combines performance, accuracy, ease-of-use, and speed for the analysis of finished wine, must and must under fermentation samples in less than 45 seconds—all in one compact solution.

The SpectraStar™ XT-R Benchtop NIR can analyze virtually any sample type in less than a minute, including solid, slurry, or liquid samples.

AZoM.com - An AZoNetwork Site

Owned and operated by AZoNetwork, © 2000-2022